[docs]def polcal(vis, caltable='', field='', spw='', intent='', selectdata=True, timerange='', uvrange='', antenna='', scan='', observation='', msselect='', solint='inf', combine='obs,scan', preavg=300.0, refant='', minblperant=4, minsnr=3.0, poltype='D+QU', smodel=[''], append=False, docallib=False, callib='', gaintable=[''], gainfield=[''], interp=[''], spwmap=['']):
r"""
Determine instrumental polarization calibrations
[`Description`_] [`Examples`_] [`Development`_] [`Details`_]
Parameters
- vis_ (string) - Name of input visibility file
- caltable_ (string='') - Name of output gain calibration table
- field_ (string='') - Select field using field id(s) or field name(s)
- spw_ (string='') - Select spectral window/channels
- intent_ (string='') - Select observing intent
- selectdata_ (bool=True) - Other data selection parameters
.. raw:: html
<details><summary><i> selectdata = True </i></summary>
- timerange_ (string='') - Select data based on time range
- uvrange_ (variant='') - Select data within uvrange (default units meters)
- antenna_ (string='') - Select data based on antenna/baseline
- scan_ (string='') - Scan number range
- observation_ ({string, int}='') - Select by observation ID(s)
- msselect_ (string='') - Optional complex data selection (ignore for now)
.. raw:: html
</details>
- solint_ (variant='inf') - Solution interval
- combine_ (string='obs,scan') - Data axes which to combine for solve (obs, scan, spw, and/or field)
- preavg_ (double=300.0) - Pre-averaging interval (sec)
- refant_ (string='') - Reference antenna name(s)
- minblperant_ (int=4) - Minimum baselines _per antenna_ required for solve
- minsnr_ (double=3.0) - Reject solutions below this SNR
- poltype_ (string='D+QU') - Type of instrumental polarization solution (see help)
- smodel_ (doubleArray=['']) - Point source Stokes parameters for source model.
- append_ (bool=False) - Append solutions to the (existing) table
- docallib_ (bool=False) - Use callib or traditional cal apply parameters
.. raw:: html
<details><summary><i> docallib = False </i></summary>
- gaintable_ (stringArray=['']) - Gain calibration table(s) to apply
- gainfield_ (stringArray=['']) - Select a subset of calibrators from gaintable(s)
- interp_ (stringArray=['']) - Interpolation mode (in time) to use for each gaintable
- spwmap_ (intArray=['']) - Spectral window mappings to form for gaintable(s)
.. raw:: html
</details>
.. raw:: html
<details><summary><i> docallib = True </i></summary>
- callib_ (string='') - Cal Library filename
.. raw:: html
</details>
.. _Description:
Description
The **polcal** task supports solving for systematic calibration
relating to the linear and circular polarization sensitivity of
synthesis observations, namely, the instrumental polarization and
cross-hand phase.
The heuristics of polarization calibration are described in more
detail
`here <../../notebooks/synthesis_calibration.ipynb#Polarization-Calibration>`__.
.. rubric:: Common calibration solve parameters
See `"Solving for
Calibration" <../../notebooks/synthesis_calibration.ipynb#Solve-for-Calibration>`__
for more information on the task parameters **polcal** shares with
all solving tasks, including data selection, general solving
properties and arranging prior calibration. Below we describe
parameters unique to **polcal**, and those common parameters with
unique properties. Since parallactic angle is usually an important
part of instrumental polarization calibration, the **polcal** task
will implicitly turn *parang=True*, and it is not a user-setable
parameter.
.. rubric:: Polarization Calibration type: *poltype*
The *poltype* parameter supports a range of modes to accommodate a
variety of situations. Except where noted, these options are not
basis-specific.
- *'Df' -* Solve for instrumental polarization (leakage D-terms),
using the transform of a specified IQU model; requires no
parallactic angle coverage, but if the source polarization is
non-zero, the gain calibration must have the correct R-L phase
registration. (Note: this is unlikely, so just use *'D+X'* to
let the cross-hand phase registration float). This will produce
a calibration table of type D.
- *'Df+X'* - For circular basis only, solve for instrumental
polarization D-terms and the cross-hand phase correction, using
the transform of an IQU model; this mode requires at least 2
distinct parallactic angles to separate the net instrumental
polarization and the cross-hand phase (same as position angle
for circulars). This will produce a calibration table of type
D. Please note: no table of type X will be produced, so you
must follow this by a run of **polcal** with *poltype='Xf'*
(see below)
- *'Df+QU'* - For circular basis only, solve for instrumental
polarization and source Q+iU; requires at least 3 distinct
parallactic angles to separate the net instrumental
polarization from the *apparent* source Q and U. This will
produce a calibration table of type D.
- 'Xf' - Solve only for the cross-hand phase (same as position
angle for the circular basis); best to use this after getting
the D-terms from one of the above modes. Requires the
observation of a calibrator with known Q+iU (or at least known
U/Q). This will produce a calibration table of type X.
- '*Dflls'* - A specialized mode which explicitly solves a
linearized form of the cross-hand data for the D-terms. *
*
- '*Xfparang+QU*' - For observations of a distinctly linearly
polarized calibrator over a significant parallactic angle
range, solve for the apparent cross-hand phase spectrum
(channelized) and integrated (per-spw) fractional linear
polarization, Q and U. This mode is intended for use prior to
solving for the instrumental polarization (e.g., Df) and works
for either the linear or circular basis, as long as there is
adequate parallactic angle to isolate the polarized calibrator
signal and cross-hand phase in the visibility cross-hands from
any instrumental polarization contribution. A significantly
polarized (> few %) calibrator should be used, and a non-zero
polarized model prior should be specified in the smodel
parameter. For the linear basis, the prior model (if reasonably
accurate) will be used to correctly resolve the 180 degree
ambiguity in the cross-hand phase and Q,U sign. For the
circular basis, an accurate prior polarized model will achieve
a good nominal position angle calibration; if the polarization
model is specified without accurate prior position angle
knowledge, the resulting Q,U estimate will be relative, and a
traditional position angle calibration will be necessary (after
solving for the instrumental polarization).
- *'PosAng'* - Solve directly for the absolute position angle
offset spectrum, for either the linear or circular basis. The
resulting calibration table stores the position angle offset
directly, as an angle in radians. Note that for the circular
basis (only!), this is interchangeable with the position angle
calibration traditionally provided by general cross-hand phase
calibration (e.g., poltype='Xf', for which position angle is
0.5 times the cross-hand phase). For the linear basis, this is
the only way to directly calibrate the absolute position angle.
.. warning:: 'Xfparang+QU' and 'PosAng' are considered "experimental" in
CASA 5.5. As more experience is gained in their use,
additional advice will be added here.
In each of these options, the *'f'* causes channelized solutions
to be obtained, which is usually desirable for modern instruments.
Omitting the *'f'* will generate strictly unchannelized solutions.
.. _Examples:
Examples
Please click
`here <../../notebooks/synthesis_calibration.ipynb#Polarization-Calibration>`__
for some **polcal** examples.
.. _Development:
Development
No additional development details
.. _Details:
Parameter Details
Detailed descriptions of each function parameter
.. _vis:
| ``vis (string)`` - Name of input visibility file
.. _caltable:
| ``caltable (string='')`` - Name of output gain calibration table
.. _field:
| ``field (string='')`` - Select field using field id(s) or field name(s)
.. _spw:
| ``spw (string='')`` - Select spectral window/channels
.. _intent:
| ``intent (string='')`` - Select observing intent
.. _selectdata:
| ``selectdata (bool=True)`` - Other data selection parameters
.. _timerange:
| ``timerange (string='')`` - Select data based on time range
.. _uvrange:
| ``uvrange (variant='')`` - Select data within uvrange (default units meters)
.. _antenna:
| ``antenna (string='')`` - Select data based on antenna/baseline
.. _scan:
| ``scan (string='')`` - Scan number range
.. _observation:
| ``observation ({string, int}='')`` - Select by observation ID(s)
.. _msselect:
| ``msselect (string='')`` - Optional complex data selection (ignore for now)
.. _solint:
| ``solint (variant='inf')`` - Solution interval
.. _combine:
| ``combine (string='obs,scan')`` - Data axes which to combine for solve (obs, scan, spw, and/or field)
.. _preavg:
| ``preavg (double=300.0)`` - Pre-averaging interval (sec)
.. _refant:
| ``refant (string='')`` - Reference antenna name(s)
.. _minblperant:
| ``minblperant (int=4)`` - Minimum baselines _per antenna_ required for solve
.. _minsnr:
| ``minsnr (double=3.0)`` - Reject solutions below this SNR
.. _poltype:
| ``poltype (string='D+QU')`` - Type of instrumental polarization solution (see help)
.. _smodel:
| ``smodel (doubleArray=[''])`` - Point source Stokes parameters for source model.
.. _append:
| ``append (bool=False)`` - Append solutions to the (existing) table
.. _docallib:
| ``docallib (bool=False)`` - Use callib or traditional cal apply parameters
.. _callib:
| ``callib (string='')`` - Cal Library filename
.. _gaintable:
| ``gaintable (stringArray=[''])`` - Gain calibration table(s) to apply
.. _gainfield:
| ``gainfield (stringArray=[''])`` - Select a subset of calibrators from gaintable(s)
.. _interp:
| ``interp (stringArray=[''])`` - Interpolation mode (in time) to use for each gaintable
.. _spwmap:
| ``spwmap (intArray=[''])`` - Spectral window mappings to form for gaintable(s)
| Only used if callib=False
| default: [] (apply solutions from each calibration spw to
| the same MS spw only)
| Any available calibration spw can be mechanically mapped to any
| MS spw.
| Examples:
| spwmap=[0,0,1,1] means apply calibration
| from cal spw = 0 to MS spw 0,1 and cal spw 1 to MS spws 2,3.
| spwmap=[[0,0,1,1],[0,1,0,1]] (use a list of lists for multiple
| gaintables)
"""
pass